Tissue Tek Manual

The TIssue-Tek'VlpTM 5Vacuum Infiltration Processor is an automatic, self-contained tissue processor, which holds up to 150 tissue cassettes or up to 300 cassettes, as seiected by the user. The VlpTM 5 is available as a floor model (Figure 1-1)and bench top model (Figure 1-2). The VIP software is programmable for up to twenty. Warranty Card (located inside the operating manual (USA only) The Tissue-Tek VIP 6 AI is a precision instrument and must USB Memory P/N A3-60-4499 be handled accordingly. Rough handling or dropping the Power Cord P/N A4-01-0542 instrument will disturb or damage internal components. Chart filters excel mac. Page 16: Positioning The Instrument 5. The Tissue-Tek TEC 5 Embedding Centre provides a compact, single workstation that enables both specimen orientation and paraffin embedding.

Tissue Tek 5 Manual
- Videos
- News
DESCRIPTION
Tissue processor.
Designed for streamlined operation, the VIP 6 utilizes enhanced reagent features for greater productivity including automatic transfer during processing, illuminated cabinet that enables observation of reagent levels, and 7 mixing options. Additionally, VIP 6 features smart instruments designed to make processing easier than before. Continuous mixing, short cycles, and a heated retort lid take processing to the next level. With VIP 6, you can introduce a safe, secure, and environmentally friendly system into your lab. Solution manager software safeguards all steps taken during processing to protect tissues and specimens. VIP 6 is the choice you know for ultimate efficiency and reliability that you expect.
FORUMSView All (14)
Ask a New Question0Repliesa month ago | a month agoSakura Tissue Tek VIP 6 AI manual mode I have seen this in two of the new AI versions of the Sakura Tissue Tek VIP 6 AI: In manual mode, when the rotary valve is at a wax station or at the Xylene Cleaning station (station 16), it doesn't seem to want to allow a pump-in. I waited while it heated the retort and the rotary valve and gate valve - it appeared to heat both valves to 70C to 71C, and the retort to the proper temperature (~63C), but wouldn't let me initiate a pump in, no matter how long I waited. In programmed mode, or in a Clean Cycle, there was no problem. Has anyone else seen this? Also, does anyone have a PM form newer than version 1 from 2012 (especially one for the AI version) and/or a service manual (new enough to cover the AI version) that they can submit? Reply |
-Ross a year ago | a year agoSakura Tissue Tek VIP 6 - ultrasound level sensor Sakura Tissue Tek VIP 6 - ultrasound level sensor fail (EC 322 & 922). I've got a machine with an intermittently failing lower level sensor. I removed the sensor to clean it more thoroughly - it looks like bright clean stainless steel. The recommendation on the Sakura PM checklist is to use acetic acid or Metaflux, but I didn't have either on hand. Sakura doesn't specify which Metaflux product, but I've never run into this issue before. Has anyone else had a level sensor issue which was corrected by intensive cleaning? Does anyone have a source for these level sensors, other than Sakura? Reply |
1Replya year ago | a year agoError 230 Can I still run my processor if the error 230 popped up during the clean retort?Reply |
Tissue Tek Glas Manual
DOCUMENTS / MANUALSView All
FEATURES
- Optional xylene-free processing
- Automatically transfers 3 reagents at a time including paraffin during processing
- Continuous mixing for efficient processing
- High-resolution LCD color touch screen allows easy access to programs
- 3 user-defined cleaning cycles with optional drying cycle Programmable rinses with water, alcohol, water/alcohol, alcohol/water, or an external solution
- Options for user-defined solutions, solution configurations, and programs
- Advanced filtering combines fume condenser system and 2 disposable activated carbon filters0014eliminates the use of water
- User-defined 150- or 300-cassette capacity configuration
- Two levels of password protection for complete instrument and information security
SPECIFICATIONS
Height | 54 in |
Length | 27 in |
Width | 24 in |
Additional Specifications

Tissue-tek Manual Slide Staining Set
Process Capacity: Up to 300 Tissue-Tek® Uni-Cassettes®
Fume Control: Fume condenser system and 2 disposable activated carbon filters
Reagent Reservoirs: 10 reagent reservoirs, 4 removable paraffin reservoirs
Program Memory: 50 processing programs, 3 user-defined cleaning programs
Password Protection: 2 levels, 4 managers, and 20 users